Medical inspection equipment – ECG monitor

ECG







Lead selection:Full lead, I, II, III, AVF, AVR, AVL, V1 -- 6 chest lead display
Gain selection:×1 、×2 、×4、×0.5
Frequency:Diagnosis mode: 0.05 ~ 130Hz
Monitoring mode: 0.5 ~ 40Hz 
Operation mode: 1 ~ 20Hz
Common mode signal suppression:diagnostic mode > 90 dB
 Monitoring mode > 100 dB   
 Operation mode > 100 dB
Scanning speed:12.5 mm / s, 25 mm / s and 50 mm / s.
Heart rate range:adults  15-300bpm (beats / min)
Neonatal / child: 15-350 BPM (beats / min)
Heart rate accuracy: ± 1% or ± 1bpm, whichever is greater 
Calibration signal: 1mV (peak value, accuracy ± 5%)
  ST segment detection:measurement range (- 2.0MV ~ + 2.0MV)
Baseline recovery time: < 3 seconds after defibrillation 
 Alarm mode:sound and light alarm
Preset range of heart rate alarm: (alarm when exceeding preset value)The upper limit is adjustable between 40 and 255 times / min
The lower limit is adjustable between 20 and 160 times / min









Resp breathing







Respiratory measurement range:adult 7-120 brpm (times / min)
Children and newborns: 7-150 brpm (times / min)
Accuracy: +/-  2 brpm (times / min) 
 Resolution:   1 brpm (time / min)
  Gain selection:  × 0.5, × 1, × 2, × 4  
    Alarm mode:    sound and light alarm
Alarm preset range: (alarm when exceeding preset value)The upper limit is adjustable between 10 and 100 times / min
The lower limit is adjustable from 0 to 99 times / min









NIBP blood pressure







Technique:oscillography (pulse wave oscillation)
 Working mode: manual / automatic
 Measurement unit:        mmHg / kPa
Over voltage protection: double safety protection 
Adult mode: 300 mmHg 
 Child mode 240 mmHg
  Neonatal mode: 150 mmHg
Blood pressure measurement range and accuracy:adultSystolic blood pressure was 40-270mmhg  
Diastolic blood pressure was 10-215mmhg
The average pressure is 20-235 mmHg
childrenSystolic blood pressure was 40-200mmhg 
Diastolic blood pressure was 10-150 mmHg 
   The average pressure is 20-165mmhg
newbornSystolic blood pressure was 40-135mmhg
  Diastolic blood pressure was 10-100mmhg 
    The average pressure is 20-110mmhg
Static pressure range: 0 ~ 300MMHG  
 Static pressure accuracy:  +/- 3mmHg  
 Blood pressure accuracy:the maximum average error was +/- 5 mmHg;
 the maximum standard deviation was 8 mmHg 
     Alarm mode:     sound and light alarm
Alarm preset range and error: (alarm when exceeding preset value)The upper limit of systolic blood pressure is 40-250mmhg; the lower limit is 10-220mmhg 
  The upper limit of diastolic blood pressure is 20-250mmhg; the lower limit is 10-220mmhg









SpO2 blood oxygen saturation







Range: 70-100%: +/- 2 digit0% ~ 69%: not defined
Oxygen saturation measurement error:70-100%:  +/-  20% ~ 69%: not defined
Resolution:1%
 Pulse frequency:  20-300 times / min  
 Alarm mode: sound and light alarm
Alarm preset range and error: (alarm when exceeding preset value)The upper limit is 20% - 100%;
The lower limit is 10% - 99%;









Temp temperature







Technology:high sensitive thermistor probe
Temperature measurement range: 0 ℃~ 45 ℃  
 Resolution: 0.1 ℃
Temperature accuracy: +/-  0.1℃ (excluding sensor error)  
   Alarm mode:sound and light alarm
Alarm preset range and error: (alarm when exceeding preset value)The upper limit is adjustable between 20.1 ℃ and 45 ℃ 
The lower limit is adjustable between 20 ℃ and 44.9 ℃

Optional built-in lithium battery 2200mAh