Medical inspection equipment – ECG monitor GT6800


1、 Performance characteristics of multi parameter monitor:        






1. Using world famous brand module, high precision, accurate parameters, reliable quality and stable performance                   

2. High brightness and high resolution 12 inch color TFT liquid crystal display, data waveform is clear and easy to see                                                                                                            

3. Light and strong body, easy to carry                                                                                         

4. Internal all metal structure, fully isolated floating ground, parameters more stable and reliable   

5. Rotary flying edge operation mode, Chinese and English operation interface, easy to learn, easy to use, easy to operate   

6. LFG full digital light frequency blood oxygen measurement technology, pulse oxygen saturation measurement more accurate 

7. Full lead seven channel synchronous display ECG      

8. Five minute ECG waveform playback, ST segment automatic detection, arrhythmia analysis, drug concentration calculation, respiratory oxygenation diagram and other functions  

 9. Full parameter 384 hours trend data memory and dynamic waveform capture           

10. The upper and lower limits of alarm can be set at will, and the automatic sound and light double alarm can be set           

11. It is suitable for adults, children and newborns  

12. Anti defibrillation, anti high frequency electrotome interference    

13. With the function of network transmission, the cable can be directly connected with the computer or send data packets through the network 

14. Ultra quiet, fanless design     

15. Standard configuration plug-in lithium battery, both AC and DC      

16. Optional built-in dual channel thermal array recorder, real-time printing waveforms and charts           

17. It can be equipped with VGA function interface and external large screen display         

18. It is widely used in internal medicine, surgery, operating room, obstetrics and Gynecology, pediatrics, emergency center, ICU, CCU














ECG









Lead selection:Full lead, I, II, III, AVF, AVR, AVL, V1 -- 6 chest lead display

Gain selection:×1 、×2 、×4、×0.5

Frequency:Diagnosis mode: 0.05 ~ 130Hz

Monitoring mode: 0.5 ~ 40Hz 

Operation mode: 1 ~ 20Hz

Common mode signal suppression:diagnostic mode > 90 dB

 Monitoring mode > 100 dB   

 Operation mode > 100 dB

Scanning speed:12.5 mm / s, 25 mm / s and 50 mm / s.

Heart rate range:adults  15-300bpm (beats / min)

Neonatal / child: 15-350 BPM (beats / min)

Heart rate accuracy: ± 1% or ± 1bpm, whichever is greater 

Calibration signal: 1mV (peak value, accuracy ± 5%)

  ST segment detection:measurement range (- 2.0MV ~ + 2.0MV)

Baseline recovery time: < 3 seconds after defibrillation 

 Alarm mode:sound and light alarm

Preset range of heart rate alarm: (alarm when exceeding preset value)The upper limit is adjustable between 40 and 255 times / min

The lower limit is adjustable between 20 and 160 times / min














Resp breathing









Respiratory measurement range:adult 7-120 brpm (times / min)

Children and newborns: 7-150 brpm (times / min)

Accuracy: +/-  2 brpm (times / min) 

 Resolution:   1 brpm (time / min)

  Gain selection:  × 0.5, × 1, × 2, × 4  

    Alarm mode:    sound and light alarm

Alarm preset range: (alarm when exceeding preset value)The upper limit is adjustable between 10 and 100 times / min

The lower limit is adjustable from 0 to 99 times / min














NIBP blood pressure









Technique:oscillography (pulse wave oscillation)

 Working mode: manual / automatic

 Measurement unit:        mmHg / kPa

Over voltage protection: double safety protection 

Adult mode: 300 mmHg 

 Child mode 240 mmHg

  Neonatal mode: 150 mmHg

Blood pressure measurement range and accuracy:adultSystolic blood pressure was 40-270mmhg  

Diastolic blood pressure was 10-215mmhg

The average pressure is 20-235 mmHg

childrenSystolic blood pressure was 40-200mmhg 

Diastolic blood pressure was 10-150 mmHg 

   The average pressure is 20-165mmhg

newbornSystolic blood pressure was 40-135mmhg

  Diastolic blood pressure was 10-100mmhg 

    The average pressure is 20-110mmhg

Static pressure range: 0 ~ 300MMHG  

 Static pressure accuracy:  +/- 3mmHg  

 Blood pressure accuracy:the maximum average error was +/- 5 mmHg;

 the maximum standard deviation was 8 mmHg 

     Alarm mode:     sound and light alarm

Alarm preset range and error: (alarm when exceeding preset value)The upper limit of systolic blood pressure is 40-250mmhg; the lower limit is 10-220mmhg 



  The upper limit of diastolic blood pressure is 20-250mmhg; the lower limit is 10-220mmhg
















SpO2 blood oxygen saturation









Range: 70-100%: +/- 2 digit0% ~ 69%: not defined

Oxygen saturation measurement error:70-100%:  +/-  20% ~ 69%: not defined

Resolution:1%

 Pulse frequency:  20-300 times / min  

 Alarm mode: sound and light alarm

Alarm preset range and error: (alarm when exceeding preset value)The upper limit is 20% - 100%;

The lower limit is 10% - 99%;














Temp temperature









Technology:high sensitive thermistor probe

Temperature measurement range: 0 ℃~ 45 ℃  

 Resolution: 0.1 ℃

Temperature accuracy: +/-  0.1℃ (excluding sensor error)  

   Alarm mode:sound and light alarm

Alarm preset range and error: (alarm when exceeding preset value)The upper limit is adjustable between 20.1 ℃ and 45 ℃ 

The lower limit is adjustable between 20 ℃ and 44.9 ℃
Data storage            









     It can transfer the data in the machine to the computer through the network cable, or send the data package to the user platform for storage and sharing















Power Supply           









Built in battery (standard): fully sealed and maintenance free super large capacity lithium battery                                    External AC: single AC power supply is 220 V, 50 Hz, and the maximum current is 1a                                                        Working temperature: 5 ℃~ 40 ℃                                                                                                       Working humidity: ≤ 85% relative humidity, no condensation                                                                             Storage temperature: - 20 ℃~ + 55 ℃                                                                                                      Storage humidity: ≤ 95% relative humidity, no condensation                                                                                              Altitude: 86.0 kPa ~ 106.0 kPa (working state)                                                                                                                    Power consumption: no more than 70va








Monitor size and weight            Dimensions: 320mm x 280mm x 190mm            Weight of finished product: 6kg















safety standards            Comply with iec60601-1, ul2601 medical international safety standards            It meets the requirements of national and industrial standards GB9706.1-2007, gb9706.9-2008, gb9706.25-2005, yy0667-2008, yy0668-2008, yy0709-2009 and yy0784-2010.




Matching            Printer specification: dual channel thermal array recorder printing mode: real time or event triggered recording waveform and text            End tidal carbon dioxide bypass mainstream