Autoclave sterilizing equipment – horizontal circular pressure steam sterilizer WS-150、200、280YDB

Characteristics • Automatially controlled the sterilizing process, easy to operate. • With drying function, suitable for the medical dressing drying. • With overtemperature, overpressure autoprotect. • The door can not be open until the pressure in the chamber is reduced to 0.027MPa. And the unit can not be started on if the door doesn't close well. • The safety valve will be open automatically when the inner pressure over 0.24MPa and the steam would be exhausted the water tank. • Automatically cut off the power if the waterlack happened, and alarm at the meanwhile. • The chamber of the sterilizer is made of stainless steel. • With printer.

Technical Data

MODEL

WS-150YDB

WS-200YDB

WS-280YDB

WS-400YDC

Technical Data

Sterilizing chamber volume

150L
φ440×1000

200L
φ515×1000

280L
φ600×1000

400L
φ700×1100

Working pressure

0.22Mpa


Working temperature

134℃


Temperature adjust range

40℃-134℃


Time for sterilization

0-99min


Time for drying

0-99min


Heat average

≤±2℃


Power

9kw/380V 50Hz

9kw/380V 50Hz

12kw/380V 50Hz

18kw/380V 50Hz

Dimension(mm)

1400×650×1600

1400×750×1700

1520×910×1900

1430×880×1830

Exterior package dimension(mm)

1560×820×1820

1560×910×1880

1680×1080×2100

1600×1050×2100

G.W/N.W

430/340kg

436/350kg

570/462Kg




Performance and features:        
           1. Automatic control of sterilization process, no need for supervision and convenient application.     
           2. It has drying function to achieve the drying effect of sterilized articles.         
           3. Equipped with over temperature and over pressure automatic protection device.       
           4. Safety interlock device: when the door is not closed tightly, the steam cannot enter the inner main body and give an alarm. When the pressure in the sterilization room is greater than 0.027mpa, the door is interlocked and cannot be opened.        
           5. Overpressure protection: automatic pressure relief protection of safety valve when the pressure exceeds the working pressure.         
           6. Low water level protection device: it cannot be started before the water reaches the high water level, and it will automatically cut off the power supply for protection when the low water level is short of water.    
           7. Over current protection: when the current is overloaded, the power breaker will automatically cut off the power supply.     
           8. The sterilization room is made of high-quality stainless steel.      
           9. With printing function.      

     Purpose:         
           The horizontal pressure sterilizer adopts the gravity replacement method to eliminate the cold air in the inner chamber, so that the cold air content of the steam in the inner chamber is less and the sterilization is more reliable. At the same time, in the process of sterilization, the control system automatically adjusts the inlet steam and exhaust steam according to the temperature of the inner chamber. It is mainly used in hospital supply room, operating room, pharmaceutical factory, scientific research unit laboratory and other units for sterilization treatment of medical and experimental cloth, surgical instruments and apparatus.