Medical inspection equipment – 8" ECG monitor



1、 Performance characteristics of multi parameter monitor:        









1. Using world famous brand module, high precision, accurate parameters, reliable quality and stable performance                                       2. High brightness and high resolution 8 inch color TFT liquid crystal display, data waveform is clear and easy to see                                     3. Light and strong body, easy to carry                                                                                                                           4. Internal all metal structure, fully isolated floating ground, parameters more stable and reliable                                                               5. Rotary flying edge operation mode, Chinese and English operation interface, easy to learn, easy to use, easy to operate                                  6. LFG full digital light frequency blood oxygen measurement technology, pulse oxygen saturation measurement more accurate                                7. Full lead seven channel synchronous display ECG                                                                                                             8. Five minute ECG waveform playback, ST segment automatic detection, arrhythmia analysis, drug concentration calculation, respiratory oxygenation diagram and other functions                                                                                                                       9. Full parameter 384 hours trend data memory and dynamic waveform capture                                                                                     10. The upper and lower limits of alarm can be set at will, and the automatic sound and light double alarm can be set                                     11. It is suitable for adults, children and newborns                                                                                                                         12. Anti defibrillation, anti high frequency electrotome interference                                                                                         13. With the function of network transmission, the cable can be directly connected with the computer or send data packets through the network                                                                                                                                                      14. Ultra quiet, fanless design                                                                                                                                               15. Standard configuration plug-in lithium battery, both AC and DC                                                                                             16. Optional built-in dual channel thermal array recorder, real-time printing waveforms and charts                                                                           17. It can be equipped with VGA function interface and external large screen display                                                                                         18. It is widely used in internal medicine, surgery, operating room, obstetrics and Gynecology, pediatrics, emergency center, ICU, CCU






































ECG












Lead selection:Full lead, I, II, III, AVF, AVR, AVL, V1 -- 6 chest lead display




Gain selection:×1 、×2 、×4、×0.5




Frequency:Diagnosis mode: 0.05 ~ 130Hz




Monitoring mode: 0.5 ~ 40Hz 




Operation mode: 1 ~ 20Hz




Common mode signal suppression:diagnostic mode > 90 dB




 Monitoring mode > 100 dB   




 Operation mode > 100 dB




Scanning speed:12.5 mm / s, 25 mm / s and 50 mm / s.




Heart rate range:adults  15-300bpm (beats / min)




Neonatal / child: 15-350 BPM (beats / min)




Heart rate accuracy: ± 1% or ± 1bpm, whichever is greater 




Calibration signal: 1mV (peak value, accuracy ± 5%)




  ST segment detection:measurement range (- 2.0MV ~ + 2.0MV)




Baseline recovery time: < 3 seconds after defibrillation 




 Alarm mode:sound and light alarm




Preset range of heart rate alarm: (alarm when exceeding preset value)The upper limit is adjustable between 40 and 255 times / min




The lower limit is adjustable between 20 and 160 times / min




















Resp breathing












Respiratory measurement range:adult 7-120 brpm (times / min)




Children and newborns: 7-150 brpm (times / min)




Accuracy: +/-  2 brpm (times / min) 




 Resolution:   1 brpm (time / min)




  Gain selection:  × 0.5, × 1, × 2, × 4  




    Alarm mode:    sound and light alarm




Alarm preset range: (alarm when exceeding preset value)The upper limit is adjustable between 10 and 100 times / min




The lower limit is adjustable from 0 to 99 times / min




















NIBP blood pressure












Technique:oscillography (pulse wave oscillation)




 Working mode: manual / automatic




 Measurement unit:        mmHg / kPa




Over voltage protection: double safety protection 




Adult mode: 300 mmHg 




 Child mode 240 mmHg




  Neonatal mode: 150 mmHg




Blood pressure measurement range and accuracy:adultSystolic blood pressure was 40-270mmhg  




Diastolic blood pressure was 10-215mmhg




The average pressure is 20-235 mmHg




childrenSystolic blood pressure was 40-200mmhg 




Diastolic blood pressure was 10-150 mmHg 




   The average pressure is 20-165mmhg




newbornSystolic blood pressure was 40-135mmhg




  Diastolic blood pressure was 10-100mmhg 




    The average pressure is 20-110mmhg




Static pressure range: 0 ~ 300MMHG  




 Static pressure accuracy:  +/- 3mmHg  




 Blood pressure accuracy:the maximum average error was +/- 5 mmHg;




 the maximum standard deviation was 8 mmHg 




     Alarm mode:     sound and light alarm




Alarm preset range and error: (alarm when exceeding preset value)The upper limit of systolic blood pressure is 40-250mmhg; the lower limit is 10-220mmhg 









  The upper limit of diastolic blood pressure is 20-250mmhg; the lower limit is 10-220mmhg

























SpO2 blood oxygen saturation












Range: 70-100%: +/- 2 digit0% ~ 69%: not defined




Oxygen saturation measurement error:70-100%:  +/-  20% ~ 69%: not defined




Resolution:1%




 Pulse frequency:  20-300 times / min  




 Alarm mode: sound and light alarm




Alarm preset range and error: (alarm when exceeding preset value)The upper limit is 20% - 100%;




The lower limit is 10% - 99%;




















Temp temperature












Technology:high sensitive thermistor probe




Temperature measurement range: 0 ℃~ 45 ℃  




 Resolution: 0.1 ℃




Temperature accuracy: +/-  0.1℃ (excluding sensor error)  




   Alarm mode:sound and light alarm




Alarm preset range and error: (alarm when exceeding preset value)The upper limit is adjustable between 20.1 ℃ and 45 ℃ 




The lower limit is adjustable between 20 ℃ and 44.9 ℃




















data storage            












     It can transfer the data in the machine to the computer through the network cable, or send the data package to the user platform for storage and sharing























Power Supply           












Built in battery (standard): fully sealed and maintenance free super large capacity lithium battery                                    External AC: single AC power supply is 220 V, 50 Hz, and the maximum current is 1a                                                        Working temperature: 5 ℃~ 40 ℃                                                                                                       Working humidity: ≤ 85% relative humidity, no condensation                                                                             Storage temperature: - 20 ℃~ + 55 ℃                                                                                                      Storage humidity: ≤ 95% relative humidity, no condensation                                                                                              Altitude: 86.0 kPa ~ 106.0 kPa (working state)                                                                                                                    Power consumption: no more than 70va
















































Monitor size and weight            Dimensions: 320mm x 280mm x 190mm            Weight of finished product: 6kg




















safety standards            Comply with iec60601-1, ul2601 medical international safety standards            It meets the requirements of national and industrial standards GB9706.1-2007, gb9706.9-2008, gb9706.25-2005, yy0667-2008, yy0668-2008, yy0709-2009 and yy0784-2010.
































Matching            End tidal carbon dioxide bypass mainstream